Overview

Product name: Mouse CCL3L1(Chemokine C-C-Motif Ligand 3 Like Protein 1) ELISA Kit
Reactivity: Mouse
Alternative Names: CCL3-L1; LD78; SCYA3L; SCYA3L1; MIP1AP; MIP1-Alpha-P; Small Inducible Cytokine A3-Like 1; Macrophage Inflammatory Protein 1-Alpha-P; Tonsillar Lymphocyte LD78 Beta
Assay Type: Sandwich
Sensitivity: 35 pg/mL
Standard: 5000 pg/mL
Detection Range: 78.13-5000 pg/mL
Sample Type: Serum, plasma and other biological fluids.
Assay Length: 3.5h
Research Area: Cytokine;Infection immunity;Immune molecule;
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Mouse CCL3L1. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Mouse CCL3L1. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Mouse CCL3L1, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Mouse CCL3L1 in the samples is then determined by comparing the OD of the samples to the standard curve.

標(biāo)準(zhǔn)曲線

Concentration (pg/mL) OD Corrected OD
5000.00 2.201 2.092
2500.00 1.602 1.493
1250.00 1.194 1.085
625.00 0.831 0.722
312.50 0.582 0.473
156.25 0.364 0.255
78.13 0.236 0.127
0.00 0.109 0.000

精密度

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

回收率

Matrices listed below were spiked with certain level of recombinant CCL3L1 and the recovery rates were calculated by comparing the measured value to the expected amount of CCL3L1 in samples.
Matrix Recovery range Average
serum(n=5) 83-97% 90%
EDTA plasma(n=5) 87-95% 91%
Heparin plasma(n=5) 81-95% 88%

線性

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of CCL3L1 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 85-94% 87-101% 86-97% 88-95%
EDTA plasma(n=5) 92-101% 87-98% 85-92% 79-96%
Heparin plasma(n=5) 86-95% 85-96% 89-100% 85-96%