免费久久无码精品一区二区,国产AV一区二区最新精品无删减,欧美久久男人的天堂,免费国产小视频国产

產(chǎn)品中心

大鼠纖維蛋白原(FG)酶聯(lián)免疫吸附檢測試劑盒
ELK1426
規(guī)格: 價格:
48T ¥1960.00
96T ¥2800.00

Overview

Product name: Rat FG(Fibrinogen) ELISA Kit
Reactivity: Rat
Alternative Names: F1; FI; Fbg; Coagulation Factor I
Assay Type: Sandwich
Sensitivity: 4.9 ng/mL
Standard: 800 ng/mL
Detection Range: 12.5-800 ng/mL
Sample Type: plasma
Assay Length: 3.5h
Research Area: Cardiovascular
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Rat FG. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Rat FG. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Rat FG, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Rat FG in the samples is then determined by comparing the OD of the samples to the standard curve.

標(biāo)準(zhǔn)曲線

Concentration (ng/mL) OD Corrected OD
800.00 1.946 1.842
400.00 1.528 1.424
200.00 1.107 1.003
100.00 0.821 0.717
50.00 0.535 0.431
25.00 0.342 0.238
12.50 0.228 0.124
0.00 0.104 0.000

精密度

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

回收率

Matrices listed below were spiked with certain level of recombinant FG and the recovery rates were calculated by comparing the measured value to the expected amount of FG in samples.
Matrix Recovery range Average
serum(n=5) 82-94% 88%
EDTA plasma(n=5) 84-98% 91%
Heparin plasma(n=5) 80-98% 89%

線性

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of FG and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 89-99% 83-99% 98-105% 96-101%
EDTA plasma(n=5) 85-100% 84-99% 85-97% 88-102%
Heparin plasma(n=5) 86-93% 91-101% 87-98% 95-104%