免费久久无码精品一区二区,国产AV一区二区最新精品无删减,欧美久久男人的天堂,免费国产小视频国产

產(chǎn)品中心

小鼠淀粉樣蛋白β1-40(Ab1-40)酶聯(lián)免疫吸附檢測試劑盒
ELK1493
規(guī)格: 價格:
48T ¥1960.00
96T ¥2800.00

Overview

Product name: Mouse Ab1-40(Amyloid Beta Peptide 1-40) ELISA Kit
Reactivity: Mouse
Alternative Names: Amyloid Beta Peptide 1-40
Assay Type: Sandwich
Sensitivity: 5.15 pg/mL
Standard: 1000 pg/mL
Detection Range: 15.63-1000 pg/mL
Sample Type: serum, plasma and other biological fluids
Assay Length: 3.5h
Research Area: Infection immunity;Neuro science;
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Mouse Ab1-40. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Mouse Ab1-40. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Mouse Ab1-40, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Mouse Ab1-40 in the samples is then determined by comparing the OD of the samples to the standard curve.

標(biāo)準(zhǔn)曲線

Concentration (pg/mL) OD Corrected OD
1000.00 2.095 2.003
500.00 1.573 1.481
250.00 0.987 0.895
125.00 0.799 0.707
62.50 0.532 0.440
31.25 0.389 0.297
15.63 0.211 0.119
0.00 0.092 0.000

精密度

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

回收率

Matrices listed below were spiked with certain level of recombinant Ab1-40 and the recovery rates were calculated by comparing the measured value to the expected amount of Ab1-40 in samples.
Matrix Recovery range Average
serum(n=5) 82-95% 88%
EDTA plasma(n=5) 90-105% 97%
Heparin plasma(n=5) 87-99% 93%

線性

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Ab1-40 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 80-92% 86-102% 87-103% 85-101%
EDTA plasma(n=5) 82-93% 96-105% 87-98% 85-97%
Heparin plasma(n=5) 95-104% 82-93% 81-96% 83-94%