Overview

Product name: Mouse F2(Coagulation Factor II) ELISA Kit
Reactivity: Mouse
Alternative Names: FII; TM; PT; Thrombin; Prothrombin; Pro-Thrombin; Activation peptide fragment 1; Activation peptide fragment 2
Assay Type: Sandwich
Sensitivity: 0.59 ng/mL
Standard: 100 ng/mL
Detection Range: 1.57-100 ng/mL
Sample Type: Plasma.
Assay Length: 3.5h
Research Area: Signal transduction;Hematology;
Uniprot ID: P19221
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Mouse F2. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Mouse F2. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Mouse F2, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Mouse F2 in the samples is then determined by comparing the OD of the samples to the standard curve.

標(biāo)準(zhǔn)曲線

Concentration (ng/mL) OD Corrected OD
100.00 2.103 1.998
50.00 1.652 1.547
25.00 1.106 1.001
12.50 0.812 0.707
6.25 0.525 0.420
3.13 0.303 0.198
1.57 0.214 0.109
0.00 0.105 0.000

精密度

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

回收率

Matrices listed below were spiked with certain level of recombinant F2 and the recovery rates were calculated by comparing the measured value to the expected amount of F2 in samples.
Matrix Recovery range Average
serum(n=5) 87-99% 93%
EDTA plasma(n=5) 78-92% 85%
Heparin plasma(n=5) 92-107% 99%

線性

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of F2 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 86-97% 83-96% 91-103% 96-105%
EDTA plasma(n=5) 86-93% 91-105% 87-98% 89-94%
Heparin plasma(n=5) 81-94% 96-105% 92-103% 82-96%