免费久久无码精品一区二区,国产AV一区二区最新精品无删减,欧美久久男人的天堂,免费国产小视频国产

產(chǎn)品中心

大鼠凝血因子Ⅹ(F10)酶聯(lián)免疫吸附檢測試劑盒
ELK1558
規(guī)格: 價(jià)格:
48T ¥1960.00
96T ¥2800.00

Overview

Product name: Rat F10(Coagulation Factor X) ELISA Kit
Reactivity: Rat
Alternative Names: FX; FXA; Stuart Prower Factor; Thrombokinase
Assay Type: Sandwich
Sensitivity: 14.2 pg/mL
Standard: 2000 pg/mL
Detection Range: 31.25-2000 pg/mL
Sample Type: Plasma.
Assay Length: 3.5h
Research Area: Signal transduction;Hematology;
Uniprot ID: Q63207
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Rat F10. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Rat F10. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Rat F10, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Rat F10 in the samples is then determined by comparing the OD of the samples to the standard curve.

標(biāo)準(zhǔn)曲線

Concentration (pg/mL) OD Corrected OD
2000.00 2.274 2.189
1000.00 1.697 1.612
500.00 1.194 1.109
250.00 0.831 0.746
125.00 0.527 0.442
62.50 0.361 0.276
31.25 0.230 0.145
0.00 0.085 0.000

精密度

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

回收率

Matrices listed below were spiked with certain level of recombinant F10 and the recovery rates were calculated by comparing the measured value to the expected amount of F10 in samples.
Matrix Recovery range Average
serum(n=5) 87-99% 93%
EDTA plasma(n=5) 92-107% 99%
Heparin plasma(n=5) 93-107% 100%

線性

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of F10 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 85-94% 97-103% 82-90% 92-101%
EDTA plasma(n=5) 85-92% 90-96% 79-102% 83-99%
Heparin plasma(n=5) 88-101% 86-92% 83-98% 96-103%